一种简易医用硬性内窥镜相对百分畸变检测系统

Simple medical rigid endoscopic relative percent distortion detection system

  • 摘要: 医用硬性内窥镜作为诊治医疗器械,其畸变直接影响医生对手术位置判断的准确性,因此内窥镜相对百分畸变的测量是一个亟需解决的问题。设计了一种基于图像处理的相对百分畸变检测系统,通过CCD相机采集校准靶板图像,对内窥镜视轴和目标靶板的板面进行垂直校准后,更换畸变靶板采集图像,计算机将得到的图像进行滤波、感兴趣区域提取和尺寸统计,最终获得相对百分畸变。以30°腹腔内窥镜为测量模型,实验得到设计的校准系统在待测内窥镜全视场70%位置所测量的相对百分畸变值优于未垂直校准系统2.2%。结果表明,系统的检测结果一致性较高,且准确度高于未垂直校准系统。

     

    Abstract: As a medical device for diagnosis and treatment, the distortion of medical rigid endoscopes directly affects the accuracy of the surgeons judgment of the surgical position. Therefore, the measurement of endoscopic relative percent distortion is an urgent problem to be solved. A relative percent distortion detection system based on image processing was designed, and the images of the calibration target plate were acquired by a CCD camera. After vertical calibration of endoscope visual axis and target plate, the distortion target plate was replaced to collect images. The computer filtered the obtained image, extracted the region of interest and carried out the size statistics, and the relative percent distortion was finally obtained. Using a 30° laparoscope as the measurement model, the experiments show that the relative percent distortion value measured by the designed calibration system at 70% of the full field of view of the endoscope to be measured is better than that of the system without vertical calibration by 2.2%. The results show that the system has higher consistency of detection results and higher accuracy than the system without vertical calibration.

     

/